The xenon lamp test chamber uses a xenon arc lamp that can simulate the full spectrum of sunlight to reproduce destructive light waves present in different environments, providing corresponding environmental simulation and accelerated testing for scientific research, product development, and quality control. The xenon lamp test chamber can be used for selecting new materials, improving existing materials, or evaluating durability changes after material composition changes. It can simulate the changes in materials exposed to sunlight under different environmental conditions very well. Complies with GB/T16585-1996; GB14522-93 ; GB/T16422.3-97 ; ASTM D2565
+8675526585726
13609623361@139.com
DESCRIPTION
.Product Features:
◆ Full spectrum xenon lamp.
u Multiple filtering systems to choose from.
u Water spray function.
u Relative humidity control.
u Test box air temperature control system.
u Irregular shaped flat product rack, convenient for placing products
u Xenon arc lamp tube.
The characteristics of easy installation, convenient use, and basically no need for daily maintenance.
u The service life of xenon arc lamps depends on the level of radiation illumination used. Generally, the service life of lamps is1600Hour. The lamp tube replacement is convenient and fast, and the long-lasting filter provides assurance for maintaining the required spectrum.
u The latest air-cooled technology xenon lamp solves the problem of high failure rate of water-cooled xenon lamp cooling method, and users can easily replace the lamp tube by themselves;Easy and convenient maintenance
PARAMETERS
Technical specifications and parameters:
Studio dimensions W × D × H: W500 × H600 × D500mm
Appearance dimensions W × D × H: approximately W1050 * H1850 * D1150mm
Temperature range: 0 ℃~80 ℃
Temperature deviation: ± 2 ℃ (without lighting)
Humidity range: 20-98% R.H
Humidity deviation: ± 2.5% R.H (humidity ≥ 75% R.H) ± 3% R.H (humidity ≤ 75% R.H) (without lighting)
Light source type: Imported air-cooled full solar spectral long arc xenon lamp (with an average service life of about 1600 hours)
Rainfall duration: 1 to 9999 minutes, adjustable
Rainfall cycle: 1 to 9999 minutes, adjustable interval (break)
Spectral wavelength: 290nm to 800nm
Xenon lamp power: 2 units, 3600W (lifespan: 1600 hours)
Light source wavelength range: 290-800nm
Heating power: approximately 4KW
Humidification power: about 2KW
Light source irradiance: 290~800nm: 300~550w/m2
300-400nm: 20-60w/m2
340nm: 0.3~0.6w/m2
420nm: 0.4~1.2w/m2
Distance from arc center to sample holder: 350-380mm
Lighting time: programmable and adjustable continuously
Blackboard temperature: 0 ℃~110 ℃± 2 ℃
Refrigeration method: mechanical refrigeration and air cooling
Total power and power supply: approximately 6KW; 380V ± 10% 50HZ
The main technical parameters of the above products refer to the temperature and humidity measured after two hours of constant temperature and humidity at room temperature below 26 ℃ and relative humidity ≤ 85% R.H.
3. Temperature and humidity operation control system:
Imported programmable full touch 7-inch controller
Accuracy range setting accuracy: temperature ± 0.1 ℃, humidity ± 2% R.H, indication accuracy: temperature ± 0.1 ℃, humidity ± 1% R.H
Platinum resistor for temperature and humidity sensor PT100Ω/MV
Blackboard temperature bimetallic blackboard thermometer
The humidification water supply system adopts automatic control, and the xenon lamp cooling circulating water is used
Circulating system temperature resistant and low-noise air conditioning motor Multi blade centrifugal wind turbine
The control system mainly consists of an operation panel, a central controller, various key switches, status indicator lights, and execution appliances.