The non-invasive blood pressure lifespan testing system, controlled by a computer system, can automatically cycle testing, with a short testing cycle and suitable for all types of non-invasive automatic measurement electronic blood pressure monitors. Suitable for medical device testing institutions to register and inspect non-invasive blood pressure modules such as electronic blood pressure monitors and multi parameter monitors, and should comply with the stress testing requirements of 4.7.3.3 in YY0670-2008.
+8675526585726
sales@damslab.com
DESCRIPTION
Non invasive blood pressure lifespan testing system
Product Introduction The non-invasive blood pressure lifespan testing system, controlled by a computer system, can automatically cycle testing, with a short testing cycle and suitable for all types of non-invasive automatic measurement electronic blood pressure monitors. Suitable for medical device testing institutions to register and inspect non-invasive blood pressure modules of electronic blood pressure monitors, multi parameter monitors, etc,Should comply withYY0670-2008Chinese4.7.3.3The requirements for stress testing.
1Technical Requirements:The equipment has passed at least10000After the second full-scale cycle, the safety and performance requirements in the standard should still be met. Full range cycle refers to the pressure from2.67kPa(20 mmHg)Or lower it to the pressure value, and then lower it to2.67kPa(20 mmHg)Or lower.
2、 Cuff interface/Structural requirements:after passing through1000Secondary opening and closing cycle10000time40kPa(300 mmHg)After the pressure cycle, the closure and sealing of the cuff and integrated airbag should still be intact enough to meet the other requirements of this standard.
IIBuilt in sensor pressure accuracy±0.3kPa;
·The pressure value and number of cyclic tests can be set according to the actual situation;·Equipped with pressure calibration function, segmented testing function, and power-off data protection function;·Compliant withYY0670-2008In the "Non invasive Automatic Blood Pressure Monitor"4.3Requirements for lifespan
4、 Performance parameters·The mainframe of the detector includes a control module, a communication module, and an air circuit module; ·Blood pressure monitor lifespan detection device, including the detection device host and computer system;·Test scope: one-way valve, air pump, pressure sensor, air release valve, and detection port;·Intelligent, efficient, and applicable lifespan detection device for all types of non-invasive automatic blood pressure monitors.