sales@damslab.com
+8675526585726
Load pad
Load pad

Load pad

Description:

Product Overview: The load pad complies with IEC60601-2-52 Figure 201.104 and is used for testing ME equipment for medical devices. Its diameter is (355+/-5) millimeters, and the surface of the impactor has a convex spherical curvature with a leading edge radius of 20 millimeters and a radius of 800 millimeters. It weighs 21 kilograms.

Telephone

+8675526585726

Email

sales@damslab.com

DESCRIPTION

Product Overview:

Load pad meets the requirementsIEC60601252image201.104Used for testing medical devicesMEEquipment, diameter is(355+/5)Millimeters, the surface of the impactor has a leading edge radius20Millimeter, radius800Millimeter convex spherical curvature, heavy21KG.