GB9706.1-2020 “Medical Electrical Equipment Part 1: General Requirements for Basic Safety and Basic Performance“ If both of the following conditions are met simultaneously, the pressure vessel should be able to withstand the hydrostatic test pressure: Complies with clause 9.7.5 of GB9706.1-2020 “Medical Electrical Equipment Part 1: General Requirements for Basic Safety and Basic Performance“ standard;
+8675526585726
sales@damslab.com
DESCRIPTION
GB9706.1-2020Medical electrical equipment - Part 1: General requirements for basic safety and basic performance
If both of the following conditions are met simultaneously, the pressure vessel should be able to withstand the hydrostatic test pressure:
Complies with clause 9.7.5 of GB9706.1-2020 "Medical Electrical Equipment Part 1: General Requirements for Basic Safety and Basic Performance" standard;
2、 Test requirements:
-Pressure greater than 50kPa; and
-The product of pressure and volume of a pressure vessel is greater than 200 kPa · 1
Use the following tests to verify compliance with requirements:
The hydrostatic test pressure should be the maximum allowable working pressure multiplied by a coefficient obtained from Figure 32.
Gradually increase the pressure to the specified test value and maintain this value for 1 minute.