The HIC ground impact attenuation performance tester is used to simulate and detect the impact performance of the ground surface on the head after a fall. It can accurately measure the HIC value of the head impact on the sports field and automatically calculate the strict height for safe falls on sports fields and children's playgrounds. The product does not require external force to move the heavy hammer, and the instrument can move flexibly as a whole and automatically display the detection results. 2、 Meets BS EN 1177-2008 (impact damping amusement park pavement) The Annex A test in the standard for determining critical drop height requires an impact testing device
+8675526585726
sales@damslab.com
DESCRIPTION
1、 Equipment Overview:
HICThe ground impact attenuation performance tester for sports fields is used to simulate and detect the impact performance of the ground surface on the head after a fall, and can accurately measure the impact on the head of the sports fieldHICMeasure the values and automatically calculate the strict height for safe falls on sports fields and children's playgrounds. The product does not require external force to move the heavy hammer, and the instrument can move flexibly as a whole and automatically display the detection results.
2、 Meet the standards:
1)SatisfyBS EN 1177-2008(Impact damping amusement park pavement).In the standard for determining critical drop heightAnnex ATest requirements (impact testing device)
2)GB30228-2013Safety performance requirements and test methods for ground impact attenuation of sports venues)Standard No5.1.2Test requirements
Technical Specifications:
1Simulation head:
1)Weight: Quality is4.6Kg±0.05Kg, hemispherical;
2)Diameter: ¢160mm±5mm;
3)Material:6061-T6aluminium alloy;
2Measuring bandwidth:100KHz;
3Range of acceleration sensor:0~500g,±5g;
4Measurement frequency range:0.3Hz~1000HZAmplitude error: <5%.
5Free fall impact height: not less than3m, accuracy1mm;
6Release system: Electromagnetic release (does not generate torque or any form of force that rotates the simulation head);
7Function: Measure head injury criteriaHICMeasure the maximum accelerationGmaxThe measurement shows the deceleration change curve when the head shaped falling object lands on the grounda(t).