Used to evaluate the bactericidal and antibacterial efficacy of products with antibacterial and sterilization functions against microorganisms (bacteria) in the air. Reference standards: GB/T 18801-2022 “Air Purifiers“ GB/T 18883-2022 “Indoor Air Quality Standards“ QB/T5217-2018 “Medical Environmental Air Purifiers“ GB/T 38504-2020 Evaluation Method of spray Disinfection Effect WS/T 648-2019 General Hygienic Requirements for Air Disinfectors GBT39990-2021 “Technical Conditions for Particulate Bioaerosol Samplers“
+8675526585726
13609623361@139.com
DESCRIPTION
forUsed to evaluate the effect of products with antibacterial and sterilization functions on airborne microorganisms(bacteria)The bactericidal and antibacterial efficacy.
reference standard:
GB/T 18801-2022Air Purifier
GB/T 18883-2022Indoor Air Quality StandardsQB/T5217-2018Medical Environmental Air Purifier
GB/T 38504-2020Evaluation Method of spray Disinfection EffectWS/T 648-2019General Hygiene Requirements for Air Disinfection Machines
GBT39990-2021Technical conditions for particle bioaerosol samplers
test project:microorganism(bacteria)Evaluation of Antibacterial and Antibacterial Efficacy
Cabin control system: The experimental chamber adopts Siemens12Inch touch screen control,The system is easy to operate. Integrated with various functions such as air intake and exhaust, sockets, fans, lighting, sterilization, temperature and humidity regulation, pressure control, and air purification,Comprehensive intelligent centralized control.
Main technical parameters
Size and Capacity: 2.85mX3.0mX2.4m,volume20m±0.5%
Cabin structure: Stainless steel square tube frame,The wall is made of stainless steel plate or float explosion-proof glass
Temperature control: 15~40℃,Maximum allowable error±1℃
Relative humidity control: 30%~80% RH,Maximum allowable error4%RH
Air tightness performance: airtightness:(1)pressure holding>2habove,The average hourly leakage rate should be less than0.5%When judged as qualified;(2)Using particle counting method,0.3μm/min≤800a/L
Disinfection and purification: Top UV lamp(2group30W),intensity≥450μW/c㎡(or≥450mW/m);Both intake and exhaust ducts are equipped withHEPA/ULPAHigh Efficiency Filter,filtration efficiency>99.999%
Other configurations: door(1a)Sampling port(4group)Transfer window(1a), glove hole(2a,Φ14cm),socket(inside the cabin4set/USBTheRS232each1set)
Lighting: ≥ 10001x(1groupLEDClean lamp)
Mixing fan: Fan blade diameter1.5rice, threeLeaf stainless steel
Circulating fan: air volume500㎡/h,diameter0.2m,Installation position off the ground1.5m,Away from the back wall0.4m
Target pollutants: chemical pollutants and biological pollutants
Background concentration: particulate matter0.3μm<800a/㎡;formaldehyde≤0.02mg/m,TV0C≤0.1mg/m,toluene≤0.05mg/m3
Security protection: Abnormal protection, overload protection, undervoltage and overpressure protection, and cabin negative pressure protection for temperature and humidity control units
INQUIRY
+8675526585726
13609623361@139.com